Temperature and Humidity (TH) Measurements

NSCL maintains the reference measurement standards of temperature and humidity for Syria and for the dissemination of the related units of measurement. Temperature measurements are based on the International Temperature Scale (ITS-90). The temperature standards are a Standard Platinum Resistance Thermometer (SPRT), and a Standard Thermocouple Type S (Platinum-Rhodium). The Temperature measurement services cover the full range of (- 50 to 1100) °C. The humidity standard is a two pressure humidity generator and the range covered is (10 to 95) % RH.

Measured quantity/Calibrated item Range Expanded Uncertainty (K=2) 
Standard Platinum Resistance Thermometers (SPRTs,PRTs) (- 50 to 420) °C (0.005 to 0.04) °C
Thermocouple Thermometers (TCs)  (- 50 to 1100) °C (0.5 to 3.0)  °C 
Digital Thermometers with PRTs or TCs (- 50 to 1100) °C (0.01 to 3.0) °C
Temperature Controllers with PRTs or TCs (- 50 to 1100) °C (0.1 to 3.0)  °C 
Liquid in Glass Thermometers (LGTs) (- 50 to 300) °C  (0.1  to 0.7) °C
Infrared Thermometers (IRs)   (0 to 500) °C (1.0 to 2.0) °C  
Surface Temperature (30 to 500) °C  (1 to 5) °C 
Furnaces, Liquid baths, Dry block Calibrators     (- 50 to 1100) °C (0.1 to 2) °C
Incubators, Fridges, Freezers, Autoclaves (- 50 to 50) °C (0.1 to 1) °C
Zero Temperature Point 0.000 °C (0.005 to 0.010) °C
PRTs Simulators and Indicators  ITS-Tables (0.1 to 1) °C
TCs Simulators and Indicators  ITS-Tables (0.1 to 1) °C
Thermo-hygrometers, Temperature and humidity data loggers (10 to 95) % RH (- 20 to 70) °C (0.5  to 3) % RH (0.1 to 0.5) °C
Temperature and Humidity Generators, Environmental chambers (10 to 95) % RH (- 20 to 70) °C (0.5 to 3) % RH (0.1 to 0.2 ) °C

On site services are provided for a wide variety of temperature and humidity measuring instruments. How To Request Calibration Services

 

Finding Us

  • National Standards and Calibration Laboratory
  • Masaken Barzeh
    Pre-fabricated house area
    Damascus - Syria

Contact Us